(13) Input the following contents as instrument conditions.
RACK CODE:1
SYRINGE SPEED:1
RACK PARAM:RACK CODE = 1
X1 = 57.1
X2 = 178.6
Nx = 10
Y1 = 0.0
Y2 = 148.5
Ny = 12
Z = 34
P = 5
INJ.METHOD: CUT
LEAD VOLUME: 30.0
REAR VOLUME: 30.0
ASP SYRINGE SPEED: 3
DSP SYRINGE SPEED: 2
AIR VOLUME: 2.0
NEEDLE WASH TIME SOLVENT1: 4
RINSE PORT WASH TIME: 1
SYRINGE VOL:2 (175 µL)
(14) Set the pump flow rate to 1.0 mL/min.
(15) Set the detector wavelength to 265 nm. Set the detector
time constant to 0.1 s.
(16) Set the following conditions in the data processing unit.
SAMPLE PERIOD : 100 ms
STOP TIME
: 3 min
4.1.3 Reproducibility
Check
Check the reproducibility of the injection volume in the following
procedure.
(1) Press
the
PROG
key and set the following conditions.
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1- 1 10.0 6 3.0
Pressing the
ESC
key can complete setting.
(2) Prepare 60 % methanol for the mobile phase and wash
fluid. Wait until the pump, detector, etc. are warmed up
and the baseline stabilizes.
4 - 5