3 - 60
(8) Deletion of Parameter
The parameter deletion method is shown next.
PROG
(SET PROG No.)
No. of program containing parameter to
be deleted (m)
ENT
(MODIFY=0, NEW=1)
0
ENT
(MAX) Upper limit pressure
level
ENT
(MIN)
Lower limit pressure
level
ENT
ENT
DEL ENT
ESC
1
Program number can be from 1 to 9.
2
For correction, input 0. For generating program, input 1.
3
If pump pressure rises beyond the level set here, the pump
automatically stops.
Values that can be inputted are 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Values that can be inputted are 0.0 to 60.0 MPa.
5
Press the
key until the step to be corrected appears on
the screen. If display advances beyond that step, press
the
key.
6
Move the cursor to the parameter to be deleted while
holding down ENT key.
7
To delete the parameter, press the DEL key and then
press ENT key.
8
When the ESC key is pressed, Edit is finished, and
Monitor screen appears.
2
3
4
5
1
6
7
8