2.4.2 Thermostat
(option)
(1) Temperature setup range*
1)
1 to 45 °C (in increments of 1 °C)
(2) Temperature control range*
1)
[ambient temperature – 21 °C] to [ambient tempe 25 °C]
and within the setup range
(humidity 60 %, sample vial 1.5 mL, 4 mL, 1 mL)
[ambient temperature – 15 °C] to [ambient tempe 20 °C] and
within the setup range (humidity 60 %, microplate)
(3) Temperature setting accuracy
±
3 °C
(Target temperature 4 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
sample vial (1.5 mL, 4 mL, 1 mL), without sample freezing)
±
4
°
C
(Target temperature 10 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
microplate without sample freezing)
(4) Temperature
distribution
±
2 °C
(Target temperature 4 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
and 4-corner sample vials (1.5 mL, 4 mL, 1 mL), without
sample freezing)
±
3 °C
(Target temperature 10 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
and 4-corner sample vials (each microplate), without sample
freezing)
(5) Maximum number of sample vials
120 (1.5 mL) (standard)
72 (4 mL) (option)
195 (1 mL) (option)
192 (96-holes microplate
×
2) (option)
768 (384-holes microplate
×
2) (option)
2 - 47