5.6 Discharge of Air Bubbles from Syringe
Items to be prepared:
Hypodermic syringe
A tube between a syringe and a syringe valve
(1) Remove the sealing screw from port5 of the syringe valve.
Syringe valve
Sealing screw
Fig. 5-10
Syringe Valve
(2) Connect the hypodermic syringe (furnished plastic injector)
to port4.
Hypodermic syringe
Fig. 5-11 Connecting the Hypodermic Syringe (or Tube)
5 - 12