1. OUTLINE
The liquid chromatograph system consists of a pump for feeding
a mobile phase, a sample introducing section (autosampler,
manual injector) for injecting a sample, a column for
chromatographic separation of the injected sample, a column
oven for maintaining the column at a constant temperature, and
a detector section for detecting a signal of a separated
component of the sample.
In addition, a degassing unit for deaerating an eluent, a gradient
device for varying a mixing ratio of eluents with time, and other
optional devices are incorporated in the system.
1.1 Isocratic System
The isocratic system has a simple configuration in which single-
solvent analysis is performed.
Chromaster 5440 Fluorescence
detector
Detector cell
Column
Chromaster 5210
Autosampler
Chromaster 5310 Column
oven
Mobile phase
Wash fluid
Injection valve
Wash pump
Syringe valve
Degassing
unit
Drain
Fig. 1-1 Isocratic System
(1) An eluent is fed by the pump through the degassing unit.
(2) A sample is injected from the autosampler.
(3) The injected sample is separated through the column which
is maintained at a constant temperature by the column oven.