4.3.2
4.4 Troubleshooting
4.4.1 Troubleshooting of Column Oven Main Unit
If a fault is noticed in the Chromaster 5310 column oven main
unit, take a remedy with reference to the following table.
No.
Symptom
(Error message)
Cause Remedy
1 ROM
ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE
2 RAM
ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE
3 PARAMETER
ERROR
PARAMETERS ARE
SET TO DEFAULT
VALUE
4 LOG
INFORMATION
ERROR
LOG INFORMATION IS
INITIALIZED
5 EEPROM
ERROR
INSTRUMENT INFO. IS
INITIALIZED
Erroneous operation of
instrument or circuit board
If error recurs after power is restored,
contact your nearest service office of
Hitachi High-Technologies
Corporation sales representative.
6 DATA
BUFFER
OVERFLOW
PRESS “CL” KEY TO
CLEAR MESSAGE
Error in communication
Check connection with chromato-
data station (CDS) and GUI
controller. And then try reconnection
with CDS.
7 CHECK
TIME
PROGRAM (TEMP)
PRESS “CL” KEY TO
CLEAR MESSAGE
Temperature specified in time
program exceeds upper
temperature limit of
-5
°
C.
Adjust upper temperature limit setting
or temperature setting in time
program.
8
TEMP CONTROL OFF
BY EXTERNAL ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE
Temperature control has been
stopped due to external
contact input.
Press the CL key for clearing an error
indication.
9
TEMP CONTROL OFF
BY LEAKAGE (GAS)
PRESS “CL” KEY TO
CLEAR MESSAGE
1. Solvent leaks from flow
path in column oven
2. Solvent remains in oven
3. Alarm level of gas leak
sensor is too low
1. Wipe up leaked liquid, and
retighten tube connections.
If gas leak error still occurs,
replace tubing.
2. Eliminate all of remaining solvent.
3. Adjust alarm level of the gas leak
sensor.
10
TEMP CONTROL OFF
BY LEAKAGE (LIQUID)
PRESS “CL” KEY TO
CLEAR MESSAGE
1. Solvent leaks from flow
path in column oven
2. Solvent remains in oven
3. Contamination is attached
to liquid leak sensor.
1. Wipe up leaked liquid, and
retighten tube connections.
If liquid leak error still occurs,
replace tubing.
2. Eliminate all of remaining solvent
in drain tray.
3. Wipe up contamination attached
in liquid leak sensor.
11 ABNORMAL
TEMPERATURE (OVEN)
ABNORMAL
TEMPERATURE
(PELTIER)
1. Temperature sensor
abnormal
2. Circuit board faulty
If error recurs after power is restored,
contact your nearest service office of
Hitachi High-Technologies
Corporation sales representative.
4 - 11