4.1.5
4.1.5 How to Check the Temperature Setting Accuracy
Check the accuracy of the temperature indicated by the column
oven in the procedure given below.
<Item to be prepared>
•
Hexagonal wrench (P/N: N315005, size: B4)
(1) Open the front door of the Chromaster 5310 column oven
to remove an inside cover.
(2) Remove the two screws, which fasten the preheat plate at
the center, using the furnished hexagonal wrench, and
fasten the protective tube of the temperature sensor by
placing it between the preheat section and preheat plate.
Protect tube (300 mm long)
Fig. 4-3 Temperature Sensor (sketch)
Temperature
sensor position
Preheat plate screw
Preheat plate
Piping connection opening
Overall View
4 - 5