3.4 Analyzing Operation
3.4.2 Setting of Recorder and Spectrum Output
4
RECORD
(Record)
The setup screen appears.
Input Numeral
Function
1
RCD RANGE : Sets the full scale range of
recorder.
2
RCD SPEED : Sets the spectrum output speed
of recorder.
3
SPECT RCD : Sets spectral data to be output.
(a) Setting the Full Scale Range of Recorder (“1” key followed
by ENT key)
Press [numerical] keys (within [0][.][0][0][1] to [2][.][5][0][0]),
and then the [ENT] key.
* Enterable range: 0.001 to 2.500 AU, enterable in
0.001-AU steps
(b) Setting the Spectrum Output Speed of Recorder (“2” key
followed by ENT key)
1 SPECT
RCD=3)
RCD RANGE (0.001 – 2.500) AU/FS
2.500
RCD SPEED (40 nm/min=1, 60 nm/min=2)
1
RECORD (RCD RANGE=1, RCD SPEED=2,
For 40 nm/min : Press [numerical] key ([1]), and then the
[ENT] key.
For 60 nm/min : Press [numerical] key ([2]), and then the
[ENT] key.
3 - 36