Press numeric key 2 and then the ENT key.
(vi) The
1
st
monitor screen returns.
(vii) Press the RECORD key. The recorder setting
screen reappears.
<Recorder setting screen>
RECORD (RCD RANGE=1, RCD SPEED=2,
3
SPECT
RCD=3)
Press numeric key 3 and then the ENT key.
(viii) The output data spectrum No. setting screen
appears.
<Output data spectrum No. setting screen>
RECORD SPECTRUM NO. (1 - 4)
1
Press a numeric key 1 to 4 and then the ENT key.
* Set the same number as was entered when
making the setting for spectrum measurement.
(ix) The background spectrum No. setting screen
appears.
<Background spectrum No. setting screen>
BACKGROUND SPECTRUM NO. (0 - 4)
0
Press numeric key 0 and then the ENT key.
(x) The output ready monitor screen appears.
<Output ready monitor screen>
RCD NO.
SPEED
WL-RANGE (Em)
WIDTH: STD
1 60 350-450
PRESS
“START”
Press the START key.
* A spectrum is recorded by starting the
recorder or chromato-data processor
simultaneously with pressing the START key.
4 - 8