(9) Check that the wash port is located just below the needle.
When it is not correct location, return to the procedure (2)
to readjust the wash port and fix it temporarily.
(10) When the wash port is located just below the needle, move
the needle once again (refer to (2)) and fix the washing port.
(11) After adjustment, press the
IDLE
key to set the needle to its
home position. Mount the needle cover in place.
4 - 15
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......