3. Individual Contacts of Each Unit
(1) Contact Input Terminals
These are contact signal input terminals for control using
contact signals from an external switch, relay, etc. Any of
the signals is activated by shorting the terminal for 0.1
second or longer. The input circuit is shown in Appended
Fig. 2-3.
(2) Contact Signal Output Terminals
These are contact signal output terminals for control of an
external unit using contact signals. The output circuit is as
shown in Appended Fig. 2-4. The rating of contacts used
is 12 V, 0.1 A. Make sure the load connected does not
exceed this rating.
e-Line cable
Other unit
Chromaster series unit A
Chromaster series unit B
Appended Fig. 2-1 e-Line Cable Connection
Appended Fig. 2-2 e-Line Contact Circuit Configuration
-
+
Max. 10 mA/unit
Max. 12 V (reverse
polarity prohibited)
+5 V
320
Ω
IN
OUT 4.7k
Ω
ON : 5 V
OFF :
0 V
Max. 0.1
(H)
(L)
(G)
A
APPENDIX 2 - 3
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...