3.4 Analyzing Operation
3.4 Analyzing
Operation
3.4.1 Spectrum
Measurement
The spectrum measurement mode setup screen appears.
Input Numeral
Function
1
ABS
: Stores spectral data in
memory.
2
BACK GROUND : Stores background spectrum
in memory.
3
D2 REF ENERGY : Stores the reference-side
energy spectrum with D
2
lamp
in memory.
4
W REF ENERGY : Stores the reference-side
energy spectrum with W lamp
in memory.
5
Hg
: Stores the energy spectrum
of mercury lamp.
(a) Storing the Spectral Data in Memory (“1” key followed by
ENT key)
Set the minimum wavelength of scan range.
5
SPECT
(Spectrum)
SPECTRUM MODE (ABS =1, BACK GROUND=2,
1
D2 REF =3, W REF =4, Hg =5)
SCANNING WL RANGE (WL1 WL2) nm ABS
<200 – 850>
200 – 600
Press [numerical] keys (within [2][0][0] to [8][5][0]), and then
the [ENT] key.
* Enterable range: 200 to 850 nm, enterable in 1-nm
steps
3 - 30