3.2 Utility Setting
3.2 Utility
Setting
SET UP
UTLTY
(Utility)
The utility setup screen appears.
1
USE PROG=3, POLARITY=4, OTHERS=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
Input Numeral
Function
1
OFFSET
: Set an offset.
2
TIME CONSTANT : Set a time constant.
3
USE PROG
: Select or avoid use of a time
program.
4
POLARITY
: Select either polarity.
5
OTHERS
: Other setting than above.
3.2.1 Setting the Offset
The offset is a set input signal value that is added to the actually
measured value for output.
SET UP
1)
Press key.
UTLTY
2)
The utility setup screen appears.
Utility Setup Screen
1
USE PROG=3, POLARITY=4, OTHERS=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
Select OFFSET.
Press [numerical] key [1], and then the [ENT] key.
3)
The offset setup screen appears.
Offset Setup Screen
0.000
OFFSET (0.000 – 2.000) AU
3 - 10