4.6 Error Messages
Message Description
Remedy
ROM ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE.
ROM is faulty.
1) Press CL
key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
RAM ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE.
RAM is faulty.
1) Press CL
key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
PARAMETER ERROR
PARAMETERS ARE
SET TO DEFAULT
VALUE.
Analytical parameters are
inadequate.
1) Press CL
key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
LOG INFORMATION
ERROR
LOG INFORMATION IS
INITIALIZED.
Log information is
inadequate.
1) Press CL
key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
EEPROM ERROR
INSTRUMENT INFO. IS
INITIALIZED.
Instrument parameters are
inadequate.
1) Press CL
key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
Ex (Em) SIDE OF WL
DRIVE MECHANISM
ERROR
Failure in detecting Ex or Em
wavelength origin during
self-diagnosis test
(initialization).
1) Cannot be released by key
operation.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
Ex (Em) SIDE
ACCURACY (254 nm)
NOT DETECTED
Wavelength error exceeds
±
10 nm during wavelength
accuracy check.
1) Carry out troubleshooting (4.5).
2) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
Abnormal temperature
detected
1) Press CL
key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Technologies sales
representative.
LAMP OFF BY
OVERHEAT!
PRESS “CL” KEY TO
CLEAR MESSAGE.
Xe lamp turned off to prevent
its overheating.
1) Press CL
key, and 1st monitor
screen returns.
2) Use UTLTY
key to light up the lamp
again.
LAMP OFF BY
EXTERNAL INPUT
PRESS “CL” KEY TO
TURN ON.
Lamp has been turned off
due to an external contact
input. Busy status is
assumed.
1) Release the key lock if it has been
activated.
2) Press CL
key and light up the lamp
again.
4 - 33
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......