Chromaster 5440
fluorescence detector
Detector cell
Chromaster 5310
column oven
Column
Injection valve
Chromaster 5210
autosampler
Wash pump
Drain
Degassing unit
Chromaster 5110
pump
Chromaster
organizer
Detergent
Mobile-
phase
liquid
Appended Fig. 1-11 Tubing Diagram of Isocratic System
(stacked in two blocks)
Connect the inlet tube from the mobile-phase container to
the IN port of degassing unit. Connect the OUT port of
degassing unit and the inlet of pump. Connect the purge
valve of pump to port 4 of the injection valve in the
autosampler.
Connect port 5 of the injection valve to the inlet of column.
To the outlet of column, connect the inlet tube of detector
cell. Connect the detergent container to the wash pump of
autosampler through the degassing unit.