APPENDIX 1 - 18
Appended Fig. 1-13 Tubing Diagram for Drain (when
stacking module units in two blocks)
Appended Table 1-6 Drain Tubes to be Used
Drain tube B1
P/N 893-2875 (attached to organizer)
Drain tube A1*1*2
P/N 893-2872 (attached to autosampler)
Drain tube A1*3
P/N 893-2872 (attached to organizer)
Drain tube *1
P/N 080-3090 (attached to autosampler)
Drain tube *2
P/N 080-3090 (separately available)
T-joint
P/N 810-1362 (attached to organizer)
T-joint N1
P/N 890-3493 (attached to autosampler)
I-joint
P/N 893-2876 (attached to autosampler)
L-joint
P/N L369125 (attached to organizer)
5.4 GUI Controller
The GUI controller is a touch panel type operation panel.
This operation panel alone enables the user to control the
respective modules.
For operation, refer to the instruction manual of the GUI
controller.
Chromaster
organizer
Drain tube A1*3
Drain tube B1
T-joint
L-joint
Chromaster
5110/5160 pump
Chromaster 5440
fluorescence
detector
Chromaster 5310
column oven
T-joint N1
Drain tube A1*1
Chromaster 5210/5260
autosampler
I-joint
Drain tube A1*2
Drain tube *2
Drain tube *1