(c) Selection of tolerance setting (USE TOLERANCE)
Condition for reaching the temperature ready status
can be set.
NOTES
: 1. In case the use of tolerance is avoided by
entering 0 (=NO), the instrument starts a
sequential program irrespective of the present
temperature of the thermostat.
2. In case the use of tolerance is selected by entering
1 (= YES), the instrument remains in the standby
status (does not launch a sequential program) until
the present temperature of the thermostat enters a
range of the set temperature
±
tolerance.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
THERMO
THERMO UNIT (SET TEMP=1, CALIBRATION=2,
3 TOLERANCE=3, SERIAL NO.=4)
3 ENT
USE TOLERANCE (YES=1, NO=0)
0 ENT
1
1 ENT
TOLERANCE (2-15)
°
C
5
Set the tolerance range.
5 ENT
Monitor screen reappears.