Select a channel (CH).
Press numeric key 1 or 2 and then the ENT key.
* Default value: 1
* Two channels can be selected for two detectors.
(5) Channel setting is now completed and the setup initial
screen appears.
Under DETECTOR, the selected channel setting is
indicated.
<Setup initial screen>
SETUP: DETECT THERMO CONTRAST S. NO.
CH1
NO USE
4 02AB-111
* When you press the ENT key here, the setup items
selecting screen is presented to allow reentry.
Or you can advance to LCD contrast setting.
* The selected setting will be fetched upon turning on
power supply at the next time.
(6) Turn off the power switch.
3.5.2 Thermo cell setting
It is selectable whether or not to use the optional thermo cell
(1) Turn on the power switch with the
UTLTY
SET UP
key held down.
(2) The following display will be presented.
<Setup initial screen>
SETUP: DETECT THERMO CONTRAST S. NO.
CH1
NO USE
4 02AB-111
Press the ENT key.
(3) The following display will be presented.
<Setup items selecting screen>
SETUP ITEMS (DETECTOR=1, THERMO CELL=2
2
CONTRAST=3, EXCHANGE STANDARD=4)
Select THERMO CELL.
Press numeric key 2 and then the ENT key.
3 - 47