(xi) The spectrum output screen appears, and
spectrum output begins.
<Spectrum output screen>
RCD NO.
Ex
Em
FL
SPEED
MODE
1 350
350
1000.0
60
Em
(xii) When the spectrum output is completed, the 1
st
monitor screen returns automatically.
NOTE
:
Set the following conditions for the D-2500
chromato-data processor.
ATT :
5
CHART SPEED : 60 mm/min
NOTE
: If the detector and data processor are
connected via E-Line (digital), start the
output by pressing the
ALL
START
key of the
data processor.
(3) Determination of Raman Signal Intensity
S (tangent method)
Tangent
Baseline
S
(baseline method)
Fig. 4-2 Measuring the Raman Signal Intensity
Determine the voltage value corresponding to S in the
figure. When using the D-2500 chromato-data processor,
calculation can be made using the length (L in mm) of S on
the chart.
S =
×
2
x
mV
L
162
L: Length of S (mm)
x: ATT setting of data processor used in
measurement
4 - 9