2.4 Specifications
(d) Number of steps storable
Up to 100 steps in a total of nine files
(15) GLP Functions
Recording of the changeover number of times and
exchange dates of the optional changeover valve.
(16) Other Function
Liquid leak sensor
Temperature control is stopped upon detection of liquid
leak.
(17) Options
(a) GUI
controller
(b) UI
pad
(c) Selector
valves
One of the following two kinds of optional valves is
selectable.
(i) 3-column selector valve
Model............................. RV500-100
(IDEX Health&Science
LLC, USA)
Wetted part.................... SUS316, PEEK
Pressure tightness......... 35 MPa
Solvents not allowed ..... Concentrated sulfuric
acid and concentrated
nitric acid are not
applicable.
(ii) 6-way 2-position valve
Model............................. RV700-100
(IDEX Health&Science
LLC, USA)
Wetted part.................... SUS316, PEEK
Pressure tightness......... 35 MPa
Solvents not allowed ..... Concentrated sulfuric
acid and concentrated
nitric acid are not
applicable.
(d) Column management system