3 - 10
3.1.4 ON/OFF
Setting
The pump On/Off operation is described below.
(1) Press the key to start. FLOW in LED display will be
lit in blue.
(2) Re-press the key while the pump is running.
(3) The pump will be suspended.
3.1.5 Purging
Operation
Purge execution and stop operation are described here.
When carrying out purge operation, make sure a purge valve is
in the open state. When the purge valve is closed, a pressure
limiter stops the pump.
An auto purge valve opens/closes a valve automatically in
conjunction with the PURGE key operation.
(1) Press the key to start the pump.
(2) Press the key to start purging.
(3) Re-press the key to stop purging.
The auto purge valve closes a valve automatically in conjunction
with a purge time setting.
PUMP
ON/OFF
PUMP
ON/OFF
PUMP
ON/OFF
PURGE
PURGE