Table 2-1 Name and Function of Each Terminal at Rear Panel
Classi-
fication
Name Function
Contents of
Signal
Contents of
Processing
Remarks
Output EVENT
(1 to 4)
Contact turns
on/off according
to coded setting
in EVENT of the
program.
Timer
ON: CLOSE
OFF: OPEN
Pulse
OPEN
The ON/OFF
signal can output
to each TIMER by
specifying an
ON/OFF EVENT
code at a desired
time of a time
program.
1-second CLOSE
signal can output
at each TIMER by
specifying pulse
EVENT code at a
desired time of a
time program.
EVENT No. and Event code
EVENT
No.
Signal
Pulse
OFF
ON
1 10
11
12
2 20
21
22
3 30
31
32
4 40
41
42
Specifying 00 as an EVENT code stops a
pump, and EVENT code 92 rings buzzer.
1) Becomes OFF when the pulse is output in
the ON status.
2) OFF has priority over ON.
3) Contact output specification of each
terminal
Voltage ... 30 V DC max.
Current ... 100 mA max.
PRESS
OUT
Pressure value
is output to the
exterior.
0 to 1 V DC
1 V = Approx. 70 MPa (Approx. 7106 psi)
LIMIT
OUT
In case of
Pressure Limiter
error, Error
Signal is
outputted.
Input
PUMP
ON IN
Pump ON from
external system.
OPEN
50 ms
or longer
Maximum current connectable: 10 mA
Edge judgment
PUMP
OFF IN
Pump OFF from
external system.
CLOSE
1 s
CLOSE
OPEN