3 - 46
3.3.6 Pressure
Monitor
The total number of times of the double speed section error is
indicated.
The double-speed section error indicates that the pressure fails
to return to the reference level (previous pressure) even after
double-speed control (an increase of pump motor speed) was
effected to compensate for decreased pressure. Air bubbles in
the flow path may cause the error.
The pressure monitor will be reset in the following cases.
▪
Input of PUMP ON or PUMP OFF
▪
Activation of contact
▪
Start of gradient elution
▪
Receiving of start signal from integrator in e-line configuration
▪
External start of pump
▪
Input of PURGE ON or PURGE OFF
GLP (LOGBOOK=1, PRESSURE M=2, KEY LOCK=3)
2
PRESSURE MONITOR
9999 TIMES
Select PRESSURE M
2 ENT
Monitor screen
GLP
Monitor screen
The total number of times of errors
ENT or ESC