5.1
5. REPLACEMENT PARTS
5.1 Replacement Parts and Consumables
The following table lists the replacement parts and consumables
used in this instrument. The majority of these parts (such as
lamps) have a lifetime which varies significantly with usage of the
instrument.
The service life of each part is based on the assumption that the
instrument operates 160 hours/month (8 hours/day), 1920
hours/year (240 days).
Table 5-1 Replacement Parts and Consumables
Part No.
Part Name
Estimated Service Life
Remarks
890-2430 D
2
lamp
1000 hours/6 months
890-2527
W lamp
1000 hours/6 months
893-0820
Hg lamp
100 hours/2 years
With Hg lamp holder
885-2832
Cell window
9600 hours/5 years
Material: Quartz
885-2836 Packing
7.8
×
2.4
9600 hours/5 years
Packing for cell window, material:
PTFE
892-7012
Photocell circuit board
9600 hours/5 years
(*
1
)
893-0821 Thermostatic
inlet
tube
9600 hours/5 years
Length
: 458 mm,
Inner diameter : 0.25 mm,
Outer diameter : 1.57 mm,
Material :
SUS
893-0822
Standard inlet tube
9600 hours/5 years
Length
: 458 mm,
Inner diameter : 0.25 mm,
Outer diameter : 1.57 mm,
Material :
SUS
893-0823
Drain tube
9600 hours/5 years
Length
: 2000 mm,
Inner diameter : 0.33 mm,
Outer diameter : 1.57 mm,
Material :
PTFE
893-0824
Union
9600 hours/5 years
Example use : Connection
between
connecting tube
and inlet tube
Inner diameter : 0.25 mm
893-0825
Fitting
9600 hours/5 years
Example use
: Column
outlet
connection
For outer diameter : 1.57 mm
893-0826 Connecting
pipe
(SUS)
9600 hours/5 years
Example use: Piping between
column outlet and
inlet tube
For thermostatic flow cell
Length :
350
mm,
Inner diameter : 0.25 mm,
Outer diameter : 1.57 mm,
Material :
SUS
5 - 1