3.2.2 Pump Motor Control Setting
The automatic flow rate control system (STANDARD) and the
fixed flow rate control system (FIX) are available as the pump
motor control method.
NOTE
: STANDARD is selected for ordinary use.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
2 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
MOTOR CONTROL (STANDARD=1, FIX=2)
1
Select MOTOR-CTRL
2 ENT
SET UP
UTLTY
Select STANDARD
1 ENT
Select FIX
2 ENT
The motor control method is set as STANDARD.
FIX STEP AREA (0, 30 - 300)
100
Enter FIX STEP AREA .
Input range is from 0 or 30 to 300.
ENT
Monitor screen
Monitor screen
Monitor screen