3 - 57
1
Program number can be from 1 to 9.
2
For editing, input 0. For generating a new program, input
1.
3 If pump pressure rises beyond the level set here, the pump
automatically stops.
Values that can be inputted are 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Values that can be inputted are 0.0 to 60.0 MPa.
5
Input the time point of the step you want to add.
6
Flow rate valid entries are from 0.00 to 5.000.
7
Input the event code. At the set time point, you can stop
the pump, generate an audible signal, markers, times, etc.
When properly interfaced, these signals can be output to an
external system. No display will show on the screen
unless an event is set. When an event is not set, press
only the ENT key.
8 When the ESC key is pressed, Edit is finished, and
Monitor screen appears.