4.5 Troubleshooting
(cont’d)
Symptom Cause
Judgment/Check
Remedy
1) Liquid
has
leaked.
Visual check
Wipe off the liquid that has
leaked in drain receiver.
Wipe off the liquid that has
adhered to leak sensor.
5. Leak error occurs.
2) “Contaminant”
has stuck to leak
sensor.
Visual check
Wipe off “contaminant”
from the leak sensor.
4 - 20