(ii) Rack parameter setting procedure
The rack parameter is settable in a numerical
input method or in the robotic mode where the
needle is actually moved for adjustment.
2 ENT
1 ENT
Input the code of the rack which parameter is to be set.
Select “INPUT VALUE” for numerical input or “ROBOTIC” for
the robotic mode.
1 ENT
Input the X1 coordinate.
57.1 ENT
Input the X2 coordinate.
178.6 ENT
Input an Nx value.
10 ENT
Input the Y1 coordinate.
0.0 ENT
RACK CODE (1-16)
1
OPERATION MODE (INPUT VALUE=1
1 ROBOTIC=2)
X1 (0.0-180.0) mm
57.1
X2 (0.0-180.0) mm
178.6
Nx (1-40)
10
Y1 (0.0-155.0) mm
UTILITY (SEQ=1, RACK=2, WASH PARAM=3,
1 OTERS=4
0.0
(to be continued)
3 - 26