3 - 3
Description of status indication at upper right of
monitor screen:
[GR] ........Gradient program is in execution.
[IN] ..........Fluid is delivering with the initial condition
of the programming mode.
[ -- ] .........A pump is suspended in the
programming mode.
[MN] ........Manual mode
[PG] ........Purge is in execution.
[LK] .........Keyboard on operating panel is locked.
[FL] .........Column flush is in execution.
[LC] .........Conditioning of a link program is in
execution.
Table 3-2 Status Indication
State
Input Key
Pump
ON OFF
GRADIENT START [GR]
STOP
[IN]
[ -- ]
MANUAL SET
[MN]
[MN]
PURGE [PG]
KEY LOCK
[LK]
[LK]
COLUMN FLUSH
[FL]
LINK CONDITIONING
[LC]
* The key is disabled when status indication is
[LC] or [FL]
(c) 2
nd
Monitor Screen
P-UNIT P.MAX P.MIN D.MIX P.WASH [ -- ]
MPa 20.0 0.0 USE ON 1
START
STOP
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...