2.2
Table 2-2 Function of Each Key
No.
Key
Function
Remarks
1 SET
TMP
(set temp)
Used for temperature setting.
Sets temperature and also
temperature upper limiter.
Upper limiter must be set at +5
°
C or
more above the set temperature.
2 COND
(condition setting)
Used for condition setting upon
judging that temperature is ready.
Sets two items; tolerance and wait
time.
Upon setting wait time of “0”, ready
occurs when arriving within the
tolerance range for temperature.
1. Key for indication and setting of
auxiliary function contents
(UTLTY).
•
Setting of leak sensor
•
Setting of temp analog output
•
Setting of error output (contact)
•
Setting of temperature
calibration
3 SET
UP
(utility setup)
2. Used for selecting LCD contrast,
and for specifying whether or not
to use the optional selector valve
(SET UP).
2. Clicking this together with
key
opens the SETUP screen. Hold
key down until SETUP screen
appears.
Used to select GLP information.
1. Indicates number of selections of
option selector valve, and sets
valve seal replacement date.
4
(GLP)
2. Sets the key lock.
2. Key lock is released by clicking
ESC key.
5 INSERT
(set program)
1. For selecting and setting up a time
program.
6 START
(start)
Pressing this key starts time program
execution.
7 STOP
(stop)
Pressing this key stops time program
execution.
8 VALVE
(valve)
For manual setting of the position of
the optional selector valve
Pressing the VALVE1 key on the
monitor screen causes a transition to
the selection screen (effective only in
a case where the valve is connected).
9 LOCK
OFF
(escape/lock off)
1. Returns to MONITOR screen from
numerical input status, etc. in each
mode.
2. Releases the key lock.
UTLTY
GLP
PROG
7
8
4
ESC
2 - 5