Table 4-3 Autosampler Main Unit Errors (continued)
No.
Error Message
Possible Cause
Remedy
1. The inside of syringe is
dried up.
2. Syringe capacity input
code is wrong.
3. There is an obstacle on
the moving route of
syringe (vertically).
1. Fill the syringe with fluid and
retry operation.
2. Input the correct code that
matches capacity.
3. Remove the obstacle.
20 ERROR-SYRINGE
4. The electric system
malfunctions.
4. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. The syringe valve seal has
worn out.
1. Replace the seal.
2. The inside of syringe valve
is dried up.
2. Fill the valve with fluid and
retry operation.
21 ERROR-SVALVE
3. The electric system
malfunctions.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.
1. The injection valve seal
has worn out.
1. Renew the seal.
2. The inside of injection
valve is dried up.
2. Fill the valve with fluid and
retry operation.
22 ERROR-IVALVE
3. The electric system
malfunctions.
3. Contact your nearest service
office of Hitachi High-
Technologies Corporation
sales representative.