(continued)
3 ENT
APL PARAM (SYR SPD=1, NEEDLE SPD=2,
1 SYR VOL=3, AIR VOL=4)
Select an operation condition.
1 ENT
Set a syringe speed to aspirate a sample
according to the sample viscosity.
Set a slow syringe speed for a high-viscosity
sample.
4 ENT
DSP SYRINGE SPEED (1-5)
4
Set a syringe speed to discharge a sample
according to the sample viscosity.
Set slow syringe speed for a high-viscosity
sample.
4 ENT
Relation between Syringe Capacity and Syringe Speed
Syringe
Capacity
Speed
1
Speed
2
Speed
3
Speed
4
Speed
5
70 µL
0.4 µL/s
1 µL/s
4 µL/s
10 µL/s
35 µL/s
175 µL
1 µL/s
2.5 µL/s
10 µL/s
25 µL/s 87.5 µL/s
700 µL
1 µL/s
4 µL/s
10 µL/s
20 µL/s 87.5 µL/s
Monitor screen reappears.
ASP SYRINGE SPEED (1-5)
4
(to be continued)
3 - 21