3.2.5
3.2.5 Other
Settings
Input Numeral
Function
1
PROCESSOR : Set the full scale of processor
output.
2
ERROR OUT
: Select the Enable or Disable
status of error signal output.
3
LEAK SENSOR : Select use or non-use of the leak
sensor.
4
LAMP
: Select ON/OFF status of the
D
2
/W lamp.
5
GAIN
: Select a gain so as to match the
flow cell to be used.
(a) Setting the Full Scale of Processor Output
SET UP
1) Press key.
UTLTY
2) The utility setup screen appears.
Utility Setup Screen
1
USE PROG=3, POLARITY=4, OTHERS=5)
UTILITY (OFFSET=1, TIME CONSTANT=2,
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3) The OTHERS setup screen appears.
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
OTHERS (PROCESSOR=1, ERROR OUT=2
Select PROCESSOR.
Press [numerical] key [1], and then the [ENT] key.
4) The PROCESSOR setup screen appears.
3
1.0 AU/FS =3, 2.0AU/FS =4)
PROCESSOR (0.25AU/FS=1, 0.5AU/FS =2
3 - 15