(continued)
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-120 10.0 1 5.0
Input the number of injections from each vial.
1 ENT
STEP RACK VIALS VOL(µL) INJ TIME
1 1 1-120 10.0 1 5.0
Input a cycle time (min).
5 ENT
STEP RACK VIALS VOL(µL) INJ TIME
2
Repeat input by the same method as for STEP1.
When pressing the
ESC
key, the monitor screen reappears.
ESC
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0 [C]
(b) STEP deleting method
Unnecessary steps in a sequential program can be
deleted.
STEP RACK VIALS VOL(µL) INJ TIME
3 _1 1- 1 10.0 1 5.0
Display the unnecessary step and press the
DEL
key.
STEP RACK VIALS VOL(µL) INJ TIME
- 1 1- 1 10.0 1 5.0
Press the
ENT
key.
STEP RACK VIALS VOL(µL) INJ TIME
3 1 5- 5 10.0 1 5.0
The old contents of the relevant step are deleted and those
of the next step are brought in.
NOTE
: STEP No. 1 cannot be deleted.