3.3 GLP-Related Functions
3.3.2 Logbook Setting (Valve)
When using the optional selector valve, information on the
replacement date of the valve seal (a maintenance item) and
number of valve switchings after the replacement is recorded in
logbook form.
Enter the year when valve
seal was replaced.
2004 ENT
Monitor screen
Carry out setting for the
logbook in which
maintenance information
is recorded.
2 ENT
VALVE1 SEAL CHANGE (MM DD YYYY)
<2000 - 2100> 9 1 2010
VALVE1 SEAL CHANGE (MM DD YYYY)
<1 - 12> _9 1 2010
Enter the month when v
seal was replaced.
6 ENT
alve
VALVE1 SWITCHING LAST CHANGED DATE
23 times
9/1/2010
Monitor screen
ENT or ESC key
Check the logbook information.
1 ENT
Enter the day when valve
seal was replaced.
2 ENT
VALVE1 SEAL CHANGE (MM DD YYYY)
<1 - 31> 9 _1 2010
LOGBOOK (REPORT=1, RESET=2)
2
1 ENT
GLP
GLP (LOGBOOK=1, KEY LOCK=2)
1
3 - 16