(3) Assembly of Flow Cell
(a) Assemble the flow cell into the cell holder in the
orientation shown in Fig. 4-11 and fix it with the cell
clamp.
(b) Tighten the upper and lower retaining screws 1 and 2.
View from top
Excitation light
Enlarged view of flow cell
Fig. 4-13 Assembly of Flow Cell Unit
(c) Attach the cell window and shielding plate.
(d) After assembling the flow cell in the order reverse to
disassembly, flow a liquid through the unit and make
sure there is no leakage.
NOTE
: Be careful not to scratch the flow cell surface when
assembling it.
NOTE
: Take care about the orientation of the polished
surfaces of the flow cell when putting it into the cell
holder (see Fig. 4-11). A mistake in the
orientation may preclude obtaining the proper
performance from the instrument.
NOTE
: Tighten setscrew 1 fully with hand, and then
tighten setscrew 2 about 30
°
manually or tighten
setscrew 1 further 30
°
with a wrench.
Transmitted
excitation light
(Instrument front)
Polished surfaces
(with slit)
Emission
Hg light
entrance
Flow cell
Reflection
plate
Polished surface
(without slit)
(Instrument front)
(Instrument rear)
(Instrument rear)