3 - 20
3.2.3 Purging Flow Rate Setting
Set up a purging flow rate.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
3 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
PURGE SET (MANUAL=1, A.PURGE=2)
0
Select PURGE
3 ENT
Monitor screen
SET UP
UTLTY
Monitor screen
Select MANUAL
0 ENT
Select A.PURGE
(When auto purge is connected only.)
1 ENT
A.PURGE FLOW (0.001 – 9.999)
5.000
Enter purging flow rate
ENT
Monitor screen
PURGE FLOW (0.001 – 9.999)
5.000
Enter purging flow rate
ENT
A.PURGE TIME (1 – 30) min
5
Enter purging time
ENT