(i)
Turn ON the power switch, and the 1
st
monitor
screen appears.
<1
st
monitor screen>
TIME Ex Em FL PROG LAMP [I]
0.0 250
350
0.0
1 Xe
(ii)
In the present status, pull out the flow cell unit
(see Fig. 5-2). “****” will be indicated at FL on
the 1
st
monitor screen.
<1
st
monitor screen>
TIME Ex Em FL PROG
LAMP [I]
0.0
250 350 ****
1
Xe
5 - 14