4.2 Periodic
Maintenance
Estimated periodic maintenance is based on full operation of 8
hours daily and 20 days monthly with standard assay.
No. Check
Item Frequency
Procedure
1
Liquid leak check
Daily
Refer to 4.2.1
2
Washing the flow path
As required
Refer to 4.2.2
3
Cleaning
As required
Refer to 4.2.3
4
Waste solution tube
Daily
Refer to 4.2.4
5
Liquid leak sensor
Monthly
Refer to 4.2.5
6
Drain tray
As required
Refer to 4.2.6
4.2.1 Liquid Leak Check
Visually check that liquid is not leaking through any tube.
4.2.2 Washing the Flow Path
The flow path should be washed on a periodic basis. Bubbles
will be formed if the flow path between the syringe and needle is
contaminated; these bubbles will adversely affect the analytical
reproducibility of the system.
(1) When using an aqueous detergent
At the end of measurement, wash the flow path with distilled
water. If contamination is heavy, fill the syringe with
detergent and leave it overnight. After cleaning the syringe,
rinse it with distilled water. If the instrument will not be
used for a week or loner, wash the flow path adequately
with distilled water and keep the syringe filled with distilled
water. In particular, the syringe and sample loop should be
washed carefully to remove contaminants. And prevent the
inside of flow path from drying out. If the flow path dries
out though contaminated, decontamination may become
impossible.
Table 4-1 Recommended Neutral Detergents
Part No.
Detergent
(brand)
Concentratio
n
Precaution
S264000
Extran®
20 to 50 times
Dilute the detergent with ion-
exchanged water or distilled
water when using phosphorus-
free neutral detergent (Extran
MA03, etc.).
775-1365
Rinsing liquid
Not diluted
4 - 8
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......