3.3.3 Auto Purge Valve Logbook
The total number of times of the changeover after replace of a
maintenance part for an auto purge valve (option) is indicated.
This function is enabled only when the auto purge valve is
connected. This function is used as an index of a replacement
for maintenance parts of the auto purge valve.
GLP (LOGBOOK=1, PRESSURE M=2, KEY LOCK=3)
1
LOGBOOK (PUMP SEAL=1, P.WASH=2,
3 A.VALVE=3, D.MIX=4, P.VALVE=5)
Select LOGBOOK
1 ENT
LOGBOOK (REPORT=1, RESET=2)
1
Select A.VALVE
3 ENT
Monitor screen
GLP
Monitor screen
Select RESET
2 ENT
A.V. SEAL CHANGE (MM DD YYYY)
<2000 - 2100> 9 1 2010
Reset logbook
NOTE
: Enter the maintenance part for the auto purge valve
replacing date.
Enter Month, Date, Year. And press ENT.
In other words, it can’t be reset without pressing the ENT
key. Even if the previous data is still indicated, reenter a
desired date and press ENT. The total number of times
of the changeover will be indicated as zero, and
integration will be started.
Monitor screen
The total number of times of the changeover
after replacement of a maintenance part for the
auto purge valve is indicated.
ENT or ESC
1=Displaying logbook
2=Resetting logbook
Select REPORT
1 ENT
A.VALVE LAST CHANGED DATE
99999 09/01/2010