Message Description
Remedy
LAMP OFF BY
EXTERNAL ERROR
PRESS “CL” KEY TO
TURN ON.
Lamp has been turned off due
to an external error. Busy
status is assumed.
1) Release the key lock if it has
been activated.
2) Press CL
key and light up the
lamp again.
LAMP OFF BY
LEAKAGE
PRESS “CL” KEY TO
CLEAR MESSAGE.
Liquid leakage has occurred.
Xe and Hg lamps have been
turned off. Busy status is
assumed.
1) Press CL
key, and 1st monitor
screen returns.
2) Use UTLTY
key to light up the
lamp again.
Xe LAMP ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE.
Xe lamp fails to come on or go
off. Temperature upper limiter
of Xe lamp power source is
activated.
1) Press CL
key, and 1st monitor
screen returns.
2) Wait at least 5 minutes after
turning off lamp for the lamp to
cool down, then turn on power
again or re-light the lamp.
3) Replace lamp with a new one.
Hg LAMP ERROR
PRESS “CL” KEY TO
CLEAR MESSAGE.
Hg lamp fails to come on or go
off.
1) Press CL
key, and 1st monitor
screen returns.
2) Replace the lamp with a new
one.
XE LAMP EXCEEDED
EXCHANGE
STANDARD TIME
PRESS “CL” KEY TO
CLEAR MESSAGA
The cumulative turn-on time of
xenon lamp has exceeded its
standard exchange time.
1) Press CL key, and 1st monitor
screen returns.
2) Replace the lamp with a new
one.
A/Z OVER RANGE
PRESS “CL” KEY TO
CLEAR MESSAGE.
Allowable auto zero range has
been exceeded.
1) Press CL
key, and 1st monitor
screen returns.
2) Change the wavelength.
3) Replace the solvent with another
one.
4) Clean the flow cell.
NO SPECTRA
PRESS “CL” KEY TO
CLEAR MESSAGE.
Spectrum of specified No. is not
found.
1) Press CL
key, and 1st monitor
screen returns.
2) Enter a number corresponding to
an existing spectrum.
3) Perform spectrum acquisition.
SPECTRA ARE
MISMATCHED
PRESS “CL” KEY TO
CLEAR MESSAGE.
Scan mode incorrect,
Em bandpass incorrect, fixed
wavelength incorrect,
background WL range doesn’t
include entire spectrum WL
range, background spectrum
mode incorrect, or energy
spectrum incorrect.
1) Press CL
key, and 1
st
monitor
screen returns.
2) Select correct parameters.
DATA BUFFER
OVERFLOW
PRESS “CL” KEY TO
CLEAR MESSAGE.
During data acquisition, data
transfer to the interface board
has been disabled, causing an
overflow of the data buffer.
1) Press CL
key, and 1
st
monitor
screen returns.
2) Check the connection with the
interface board.
THERMO CELL IS NOT
CONNECTED
A thermo cell is not connected
even though the thermo cell
setting is in ON state.
1) Press CL key, and 1
st
monitor
screen returns.
2) Check the thermo cell
connection.
4 - 34