3.2 UTLTY Setting
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
4 ERROR OUT=3, CALIBRATION=4)
OFFSET (
°
C)
ENT or ESC key
Monitor screen
0.5
4 ENT
Select either plus or minus
side for deviation.
1 ENT
SET +OFFSET (0.0 - 5.0)
°
C
0.5
Set an offset value for the
correction.
0.5 ENT
SELECT MODE (+OFFSET=1, -OFFSET=2)
1 OFFSET
(
°
C)=-1.0
Present setting is
indicated.
SET UP
UTLTY
NOTE
: 1. For conducting temperature calibration, be sure to
use a calibrated thermometer for measuring
temperature.
2. There is a slight difference between the set and
actual temperatures depending on the temperature
setting. In temperature calibration, it is
recommended to determine the deviation at the set
temperature that is used most frequently.
3 - 14