(7) Sample
loading
Deliver the sample into the sample loop.
The syringe discharges [vf + vi + va + vd : Feed
Injection Air Dead volume (inside
parameter)], and delivers a sample into the loop.
Buffer tube
vd :Dead volume (the instrument inside parameter)
vi :Injection volume (the sample net volume to the flow path)
vf :Feed volume (the wash fluid volume to deliver a sample to the loop)
va :Air volume (the air volume which sample is sandwiched.)
Syringe valve
Syringe
Wash pump