(6) The UI pad operation is disabled when connecting with the
chromato data station or the GUI controller (option).
3.1.3 Auto Zero adjustment
The output value should be zeroed when zero point adjustment
is required or when the data baseline deviates from the recorder
zero point.
(1) Press
the
A/Z
key.
(2) The FL indication on the 1
st
monitor screen is reset to zero
and output signal becomes zero.
<The 1
st
monitor screen>
TIME Ex
Em FL
PROG LAMP [I]
0.0 250 350 0.0
1 Xe
* Setting value range: 0 to 1000
If the range is exceeded, an error message is displayed
3.1.4 Wavelength
setting
Specify Ex (Excitation wavelength) and Em (Emission wavelength).
(1) Press
the
WL
key.
(2) The following display will be presented. Set an excitation
wavelength (Ex) (cursor appears under Ex).
<Ex wavelength setting screen>
Ex
Em
(200 - 850.0) nm
250 350
Press numeric keys (2 0 0 to 8 5 0) and then the ENT key.
* This wavelength is settable within a range from 200 to
850 nm in increments of 1 nm.
* Default value: 250 nm
* Press the ESC key or
U
key to cancel the input excitation
wavelength value. The 1
st
monitor screen returns.
* Pressing the
V
key enables you to advance to the
emission wavelength setting without changing the
excitation wavelength.
3 - 6