3. Installation Place
For Chromaster Series system installation, it is required to
provide a space of 540 mm in width, and 550 mm in depth at
least.
A space of 150 mm or more in width shall be provided on the
rear side of the instrument, and a space of 100 mm or more in
width shall be provided on each of the left and right sides of the
instrument.
Prepare a level, sturdy installation table capable of bearing a
load of at least 150 kg.
In an example of a total system arrangement, Chromaster
organizer, Chromaster 5110 pump, Chromaster 5210
autosampler, Chromaster 5310 column oven, and Chromaster
5440 fluorescence detector, stacked in two blocks, are included.
Check and prepare an installation place satisfying the space
requirements for your actual system and OA equipment to be
used in combination.
4. Installation Environment
The following environmental conditions must be satisfied.
(1) Operating Ambient Temperature
Within a range of 4 to 35
°
C. And temperature variation
should be minimum during measurement. Also, be careful
not to allow condensation to form.
(2) Operating Ambient Humidity
25 to 85% RH
(3) Atmospheric
Gas
(a) There should be adequate ventilation.
(b) Should be no acidic, alkaline or other gas that would
highly corrode metal.
(c) Should be free from organic solvent gases (especially
benzene, paint thinner) that would dissolve the painted
surfaces of the instrument.