(24) Weight
Approx. 24 kg (Approx. 29 kg with thermostat)
(25) Ambient temperature
4 to 35
°
C (non-condensing)
(26) Ambient humidity
25 to 85%
(27) Other functions
Liquid leak sensor, Door lock
(28) Power supply
24 V DC, 4 A
*1) Performance specification should be checked in the
following combination. Avoid using the non-slit septum
or the sealed plastic cap to acquire original good
reproducibility performance.
Part No.
Part Name
080-3595
1.5 mL sample vial
855-2351
Septum (Si -TEF), slit (-)
080-3704 Screw
cap
In the other cases (refer to 6.3), adjust the rack
parameters (refer to 3.3) according to the combination
used for your practical application, and check the
reproducibility performance before proceeding to
operation.
For good reproducibility, avoid using a septum repeatedly
to prevent a cause of a hole. It causes a fluctuation of
sample concentration in sample vial.
*2) Avoid rapid changes in ambient temperature, and not
exposed to drafts such as an air conditioner. These
may influence the analysis result which reproducibility
with high precision required.
2 - 46