Create a time program as indicated below for measuring the
Raman spectrum of water.
TIME
Ex
Em
BASE (A/Z=1, HOLD=2)
0.0 350
450
1
0.5
360
0.6
370
0.7
375
0.8
380
0.9
385
1.0
390
1.1
395
1.2
400
1.3
405
1.4
410
1.5
415
1.6
430
1.8
450
2.0
NOTE
: The above time program shows an
example where the minimum FL value
was obtained at 380 nm and the
maximum at 400 nm. For the top Em
of the program, set a wavelength that
includes the wavelengths where the
maximum and minimum FL values were
obtained. (For instance, if the maximum
FL value was obtained at a wavelength
of 397 nm, then the wavelength at 1.1
minutes should be set at 397 nm.)
4 - 13