3 - 40
3.3 GLP
The logbook (a pump seal, an optional part, etc.) are displayed
and reset. Additionally, a keylock is set up to prevent incorrect
input during an analysis.
3.3.1 Pump Seal Logbook
The total volume of the delivery liquid after replace of a pump
seal is indicated. This function is used as an index of a
replacement for pump seal.
GLP (LOGBOOK=1, PRESSURE M=2, KEY LOCK=3)
1
LOGBOOK (PUMP SEAL=1, P.WASH=2,
1 A.VALVE=3, D.MIX=4, P.VALVE=5)
Select LOGBOOK
1 ENT
LOGBOOK (REPORT=1, RESET=2)
1
Select PUMP SEAL
1 ENT
Monitor screen
GLP
Monitor screen
Select REPORT
1 ENT
Select RESET
2 ENT
PUMP SEAL CHANGE (MM DD YYYY)
<2000 - 2100> 9 1 2010
Reset logbook
NOTE
: Enter the pump seal replacing date.
Enter Month, Date, Year. And press ENT.
In other words, it can’t be reset without pressing the ENT key.
Even if the previous data is still indicated, reenter a desired
date, and then press ENT . The total volume of the delivery
liquid will be indicated as zero, and integration will be started.
Monitor screen
SOLVENT LAST CHANGED DATE
99999 L 09/01/2010
The total volume of the delivery liquid after pump
seal replacement displays ”L”.
ENT or ESC