Press
PROG
key.
↓
SET PROGE NO. (1 - 9)
1
Press 1 key and then the ENT key.
↓
EDIT MODE (MODIFY=0, NEW=1)
0
Press 1 key and then the ENT key.
↓
TIME Ex Em BASE PMT
0.0
1
1
Press 0 key and then the ENT key.
Press 2 5 0 keys and then the ENT key.
Press 3 5 0 keys and then the ENT key.
Press 1 key and then the ENT key.
Press 3 key and then the ENT key.
↓
TIME Ex Em BASE PMT
0.0 250
350 1 3
1
Press 1 0 keys and then the ENT key.
Press 2 7 0 keys and then the ENT key.
Press 3 7 0 keys and then the ENT key.
Press 2 key and then the ENT key.
Press 2 key and then the ENT key.
↓
TIME Ex Em BASE PMT
10.0 270 370
2 2
1
Press 2 0 keys and then the ENT key.
↓
TIME Ex Em BASE PMT
20.0
1
Press the ESC
key.
Time program setting is now finished.
↓
TIME Ex Em FL PROG
LAMP
[I]
0.0 250 350 0.0
1
Xe
Press the START key.
The time program starts and measurement is initiated.
3 - 35
Summary of Contents for CHROMASTER 5160
Page 244: ......
Page 406: ......
Page 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Page 1016: ......