3.2.3
3.2.3 Setting of Error Out Contact
Specify whether or not to output an error contact signal to
another instrument upon occurrence of an error in the column
oven. When the other instrument receives the error contract
signal, its operation is stopped.
UTILITY (LEAK SENSOR=1, ANALOG OUT=2,
3 ERROR OUT=3, CALIBRATION=4)
ERROR OUT SIGNAL (DISABLE=0, ENABLE=1)
1
3 ENT
Set whether or not to output the
error contact signal.
1 ENT
Monitor screen
SET UP
UTLTY
NOTE
: Error contact output will always be issued regardless of
the setting of ERROR OUT contacts for only leak error
occurrence.
3.2.4 Setting of Temperature Calibration
If there is a deviation between the indicated temperature of the
Chromaster 5310 column oven and the actual temperature, the
deviation can be corrected by setting for temperature calibration.
For instance, if the actual temperature measured by a calibrated
thermometer is 40.5
°
C while the indicated temperature of the
column oven is 40.0
°
C, the offset value for the deviation is +0.5,
so make the setting as illustrated below.